Description & Specifications
The ETA2010 differential viscosity detector is designed to meet the requirements of modern GPC applications. Low-noise electronics combined with high-sensitivity pressure transducers as well as superior sample path design results in an excellent detection limit and high signal linearity. The instrument was designed for ease-of-use and trouble-free operation. Many instrument functions have been automated or are accessible by just pressing a single key. Solvent change and degassing of the reference cell are done by a solenoid valve. This results in easy handling and safe operation especially if toxic solvents are used. Specification Differential Pressure Range -1.6 to +10.0 kPa Inlet Pressure Range 0 to 150 kPa Detector Volume (Sample) 15 :l Noise Level 0.2 Pa, differential pressure signal, 25/C Temperature Range room temp. up to 80/C, hi-temp. model up 150/C Digital Input Flush, Zero, Sync, Error |